PDA

View Full Version : Máy kiểm soát - đo và điều chỉnh pH mini Hanna BL 981411-0


hvn.kinhdoanh1
18-05-2016, 11:47 PM
Máy kiểm soát - đo và điều chỉnh pH mini Hanna BL 981411-0

Hãng: HANNA

Xuất xứ: Romania

Model: BL 981411-0

Nguồn điện: 12 VDC

Bảo hành: 12 tháng

Giá: Liên hệ 0977.459.400

Clarkson BL981411-0

máy đo và điều chỉnh độ ph hanna (http://thietbidohanna.com/may-do-va-dieu-chinh-do-ph-129.html)

pH Mini Controller 12V DC

Hanna Instruments BL 981411-0

Description:

BL 981411 pH controller has been designed for easy, affordable installation in tight spaces, ideal for simple and effective process

control. The unit is provided with differential pH input and can be used with any pH electrode with standard BNC connector.

Measurements are clearly displayed on the large LCD, while the status LED indicates operating mode.

thiết bị đo hanna (http://thietbidohanna.com/)

BL 981411 is also provided with a dosing relay. Selecting acid dosing, will cause the relay to activate when the measurement is higher

than the set point. If the basic dosing is selected, the relay is activated when the reading falls below the set point.

Set point adjustment (from 0 to 14 pH) and calibration procedures are easily performed with trimmers on the front panel. You can also

choose from automatic or manual dosing modes with a switch on the front panel. Manual control is particularly useful during maintenance

operations, because it permits you to enable or disable the dosing relay according to your need. An overtime control system advises you

when the relay is active too long, to help prevent overdosing.

Two models are available, with power supply input at 115/230 Vac or 12 Vdc.

Máy kiểm soát - đo và điều chỉnh pH mini Hanna BL 981411-0 (http://thietbidohanna.com/may-do-va-dieu-chinh-do-ph/may-kiem-soat-ph-mini-hanna-bl-981411-0-681.html)

Specifications:



Range

pH

0.0 to 14.0 pH



Resolution



0.1 pH



Accuracy



±0.1 pH



Calibration



manual, with offset trimmer



Dosing Relay



1, max 2A (fuse protected), 250 Vac, 30 Vdc



Dosing Selection



acid or alkaline
contact open = acid dosage, activated when reading exceeds setpoint value
contact closed = alkaline dosage; activated when reading falls below setpoint value



Setpoint



adjustable from 0.0 to 14.0 pH



Overtime



adjustable, typically from 5 to approx. 30 minutes



Input Impedance



102 Ohm



Power Supply



See Chart



Dimensions



83 x 53 x 99 mm (3.3 x 2.1 x 3.9")



Weight



See Chart





Accessories:



Solutions



HI 7004L

4.01 pH Buffer Solution
1 x 500 mL bottle



HI 7007L

7.01 pH Buffer Solution
1 x 500 mL bottle



HI 7010L

10.01 pH Buffer Solution
1 x 500 mL bottle



HI 70300L

Electrode Storage Solution
1 x 500 mL bottle



HI 7061L

General Purpose Cleaning Solution
1 x 500 mL bottle



Accessories



HI 740146

Bracket for Installation of Process Meters



Order Information:

The BL 981411 is supplied with mounting brackets, transparent cover and instructions.